Normal Recording Mode
(AC 50/60Hz, Sine wave, Input: 10% or more of the range at CH1) |
| Range |
RMS Accuracy |
| 100.0mA |
|±2.0%rdg±0.9%f.s.+ Accuracy of Sensor |
| Other ranges |
|±1.5%rdg±0.7%f.s.+ Accuracy of Sensor |
| Crest Factor |
|2.5 or less: RMS accuracy (sine) +
2%rdg + 1%f.s. |
| *Max, Min and Instant Peak
values in Normal Recording mode are just reference values: their
accuracies aren't guaranteed. |
|
Trigger Recording Mode (AC 50/60Hz sine wave) |
| Range |
Accuracy |
| 100.0mA |
|±3.5%rdg±2.2%f.s.+ Accuracy of Sensor |
| Other ranges |
|±3.0%rdg±2.0%f.s.+ Accuracy of Sensor |
|
Capture/Power Quality Analysis Recording Mode |
| Range |
Accuracy |
| 100.0mA |
|±3.0%rdg±1.7%f.s.+ Accuracy of Sensor |
| Other ranges |
|±2.5%rdg±1.5%f.s.+ Accuracy of Sensor |
|
General |
| Recording Mode |
5010: Normal, Trigger, Capture
5020: Normal, Trigger, Capture, Power Quality Analysis |
| Operating System |
Successive Approximation
(CH1 single synchronized sampling) |
| Rated max. working voltage |
AC 9.9Vrms, 14V peak value |
| Number of input channel |
3 channels |
| Measurement method |
True RMS |
| RMS measuring interval |
approx. 100ms |
| Sampling interval: |
|
| Normal/Trigger Mode |
5010/5020: approx. 1.65ms/CH |
| Capture Mode |
5010/5020: approx. 0.55ms (waveform:
at every 1.1ms) |
| P.Q.A Mode |
5020: approx. 0.55ms |
| Low battery warning |
Battery mark display (in 4 levels) |
| Over-range indication |
"OL" mark is displayed when exceeding
the measuring range |
| Auto power off |
Power-off function operates
automatically after a switch remains for 3min. (when recording is
stopped) |
| Location for use |
Indoor use, Altitude up to 2000m |
| Operating temperature & humidity range |
-10ºC~50ºC / Relative humidity 85% or
less (no condensation) |
| Battery |
DC6V: Alkaline battery (LR6)×4pcs /
External supply DC9V (Special AC Adapter) |
| Possible measurement time |
Approx. 10 days (with alkaline LR6
batteries |
| Applicable standards |
IEC61010-1:2001 CAT.III 300V Pollution
degree2
IEC61326 (EMC standard) |
| Withstand voltage |
AC 3540V (RMS 50/60Hz) / for 5 sec. |
| Dimension |
111(L)×60(W)×42(D)mm |
| Weight |
Approx. 265g |
| Accessories |
Alkaline battery LR6×4pcs, 9118
(Carrying case)
PC software "KEW LOG Soft 2", 7148 (USB cable)
Instruction manual, Quick manual, Install manual,
USB Notice sheet |
| Optional |
8146/8147/8148 (Leakage & Load current
Clamp Sensor),
8121/8122/8123 (Load current Clamp Sensor),
8309 (Voltage Sensor: only KEW 5020), 8320 (AC Adapter),
9135 (Carrying Bag), 9119 (Hard Case),
7185 (Extension cable for sensor) |